812.219 Ordinance of 26 May 2022 on In Vitro Diagnostic Medical Devices (IvDO)

812.219 Ordinanza del 4 maggio 2022 sui dispositivi medico-diagnostici in vitro (ODIv)

Art. 56

1 For class C and D devices, other than devices for performance studies, the manufacturer must draw up a summary of safety and performance.

2 This summary shall be written in a way that is clear to the intended user and, if relevant, to the patient.

3 The minimum content of the summary is governed by Article 29 paragraph 2 EU-IVDR52.

4 The draft of the summary, together with the documentation, must be submitted to the designated body involved in the conformity assessment for validation by that body.

5 The manufacturer shall publish the summary after it has been validated.

6 The manufacturer must mention on the label or instructions for use where the summary is available.

52 See the footnote to Art. 4 para. 1 let. e.

Art. 56

1 Nel caso dei dispositivi delle classi C e D, diversi dai dispositivi oggetto di studio delle prestazioni, il fabbricante redige una sintesi relativa alla sicurezza e alle prestazioni.

2 Questa sintesi deve essere redatta in modo da essere chiara per l’utilizzatore previ-sto e, se del caso, per il paziente.

3 Il contenuto minimo della sintesi è retto dall’articolo 29 paragrafo 2 UE-IVDR52.

4 La bozza della sintesi va presentata unitamente alla documentazione all’organismo designato coinvolto nella valutazione della conformità per convalida.

5 La sintesi validata è pubblicata dal fabbricante.

6 Il fabbricante indica sull’etichetta o nelle istruzioni per l’uso dove è reperibile la sintesi.

52 Cfr. nota a piè di pagina relativa all’art. 4 cpv. 1 lett. e.

 

This document is not an official publication. Only the publication of the Federal Chancellery is legally binding.
Il presente documento non è una pubblicazione ufficiale. Fa unicamente fede la pubblicazione della Cancelleria federale. Ordinanza sulle pubblicazioni ufficiali, OPubl.