812.219 Ordinance of 26 May 2022 on In Vitro Diagnostic Medical Devices (IvDO)

812.219 Ordinanza del 4 maggio 2022 sui dispositivi medico-diagnostici in vitro (ODIv)

Art. 27 Assessment

1 Swissmedic shall conduct an on-site assessment of the conformity assessment body and, if relevant, of all sub-contractors and subsidiaries.

2 If Swissmedic identifies non-compliances in the course of its assessment, it shall draw up a list of non-compliances for the applicant. Swissmedic shall set a deadline for the conformity assessment body by which the latter shall submit to Swissmedic a corrective action plan to address the non-compliances and a preventive action plan.

The plans shall indicate the root cause of the identified non-compliances and shall include a timeframe for implementation of the actions therein.

4 Swissmedic shall decide whether the proposed action is suitable and whether the timeframe for implementation is appropriate.

Art. 27 Valutazione

1 Swissmedic svolge una valutazione in loco dell’organismo di valutazione della conformità e, se pertinente, di tutti i subcontraenti e le società controllate.

2 Se nella valutazione constata delle lacune, ne redige un elenco per l’organismo di valutazione della conformità richiedente. Stabilisce il termine entro cui l’organismo deve sottoporre a Swissmedic un piano d’azione correttivo destinato a sanare le lacune e un piano d’azione preventivo.

3 Nei piani devono essere indicati i motivi principali delle lacune rilevate e un termine per l’attuazione delle azioni.

4 Swissmedic decide se le azioni proposte sono adeguate e se il termine previsto è appropriato.

 